Total Organic Carbon Analyzer 0-1000 mg/L TOC-A2000 & TOC-A5000P Scitek Global
Product Description
TOC-A2000
and TOC-A5000P total organic carbon analyzer is a highly automated
laboratory water quality analyzer. May use it to analyze high salt sample types
and has automatic dilution, purging, and acid addition capabilities. The
biggest advantage of this total organic carbon analyzer is the thorough
oxidation and digestion, which can oxidize not only low molecular weight
organic compounds that are easy to decompose but also insoluble macromolecular
organic compounds that are difficult to decompose.
Benchtop Toc Analyzer Feature:
1. Auto sample dilution, auto
acid-adding and auto gas purging (TOC-A5000P).
2. 680℃ catalytic oxidation technology with platinum catalyst,
especially for seawater test.
3. NDIR detector, high sensitivity and
stability.
4. Optional autosampler.
5. Multi-functional PC software.
6. Precise gas flow control technology
improves accuracy.
7. PID temperature controlling
technology, ensure higher accuracy.
8. Unique three-stage dehydration
technology improves drying efficiency.
9. Personalized standard curve
management provides great convenience for users.
10. Consumables management reminds users
to replace consumables in time.
11. Modular design, simplifying device
operation and maintenance.
TOC Analyzer Specifications:
|
Model
|
TOC-A2000
|
TOC-A5000P
|
|
Measurement
Method
|
High Temperture
Catalytic Combustion
|
|
Detector
|
NDIR
|
|
Analysis
Parameter
|
TC, TIC, TOC,
NPOC
|
|
Control Mode
|
PC software
controlled
|
|
Gas Requirement
|
Oxygen, purity
≥99.995%
|
|
Standard
Configuration
|
Manual liquid
injection
|
|
Optional Sampler
|
Liquid
Autosampler;
Solid Autosampler
(mandatory for solid sample);
|
|
Measurement Range
|
0-1000 mg/L, can
extend to 0-100,000 mg/L (Manually dilution)
|
0-1000 mg/L, can
extend to 0-100,000 mg/L (Automatically dilution)
|
|
Limit of
Detection
|
TC: 50
μg/L, IC: 20 μg/L
|
|
Measuring Time
|
TC: approx.
4min, IC: approx. 3min
|
|
Max. Permissible
Error
|
TOC: ± 5%, IC: ±
4%
|
|
Repeatability
|
≤3%
|
|
Injection Volume
|
TC:
100-500μL, IC: 100-2000μL
|
|
Power Requirement
|
AC110/220V,
50/60Hz
|
|
Consumption
|
700 W
|
|
Machine Shipping
Size(W×D×H)
|
780710×610 mm
|
|
Accessory Box
Shipping Size(W×D×H)
|
710×230×600 mm
|
|
Machine
N.W./G.W.
|
41/59 kg
|
45/63 kg
|
|
Accessory
N.W./G.W.
|
5/7 kg
|
6/8 kg
|
Application of TOC Analyzer:
1. Wastewater treatment plants test
incoming and outgoing water
2. Water purification plant monitoring
of river water and treated water
3. Monitoring of treated tap water,
river water and rainwater
4. Continuous monitoring of plant water,
e.g. recycled water, cooling water, boiler water, etc.
How does a toc analyzer work
A TOC (Total
Organic Carbon) analyzer works by measuring the total amount of organic carbon
in a sample. The process involves three main steps:
1. Sample Preparation: The sample is
usually acidified to remove inorganic carbon (such as carbonates) by converting
it into carbon dio×ide (CO2), leaving only the organic carbon.&l